Tuesday, February 5, 2013

CR Quality Control #4 – Contrast Evaluation


Perform this test to see if the CR’s ability to visualize scale of contrast is comparable to the initial baseline image.  You can evaluate the contrast for hard copy film, the CR QC station, and the radiologist’s reading station. 

Procedure:
  1. Erase your dedicated test IP
  2. Place a lead apron on the floor of your x-ray room (use same room that baseline image was performed in)
  3. Place test IP on lead apron
  4. Place phantom on cassette (same orientation as baseline image)
  5. Raise the x-ray tube to maximum height, center, and open collimation about 1" past edges of test IP
  6. Use the same technical factors that produced the baseline image (check monthly log for exposure factors) and make an exposure
  7. Annotate exposure indicator number and which CR reader was used
  8. Visually compare the contrast areas on the new exposure of the phantom with the contrast areas on the baseline image in your archive (*note – the brightness and contrast settings on your viewing monitors should be the same as when the baseline image was evaluated)

Evaluation:

The contrast area on the left (for my phantom) consists of two circles of Pb (lead) equivalent material.  The outer circle is 0.15 mm Pb, while the inner is 0.5 mm Pb.  




The contrast area on the right consists of an empty hole, while the inner circle is made of 0.7 mm Cu (copper).




A simple visual comparison is enough to determine a pass/fail result.  Keep a few things in mind when looking at the images:
The monitors will degrade over time, which can provide some display inconsistency.  It is advised to observe both the QC monitor and the radiologist’s monitor for variation.  If only one of the monitor’s display appears to be sub-par, while the other appears optimum, consider replacing the sub-par monitor.  However, if both monitors appear sub-par, it is more likely due to a reduced contrast performance of the reader, and a service engineer should be notified.
If you are comparing using hard-copy film, you can use a densitometer to get a precise density difference measurement from each contrast area, comparing the outer circles to the inner.


Other posts in this series:

2 comments:

  1. interesting blog. It would be great if you can provide more details about it. Thank you...


    calibration company

    ReplyDelete
  2. What kind of additional details would you like to see?

    ReplyDelete

Here's a Quick Way to Speed Up Your Xray Exams

Beginning your clinical rotation in x-ray school can be overwhelming.  Not only are you plunged into a new environment, you are expected ...